SPECIFICATIONS NI USB-6003 Low-Cost DAQ USB Device The following specifications are typical at 25 C, unless otherwise noted. For more information about the NI USB-6003, refer to the NI USB-6001/6002/6003 User Guide available at ni.com/manuals. Analog Input Number of channels Differential 4 Single-ended 8 ADC resolution 16-bit Maximum sample rate (aggregate) 100 ks/s Converter type Successive approximation AI FIFO 2,047 samples Trigger sources Software, PFI 0, PFI 1
Input range ±10 V Working voltage ±10 V Overvoltage protection Powered-on ±30 V Powered-off ±20 V Input impedance >1 GΩ Input bias current ±200 pa Absolute accuracy Typical at full scale 6 mv Maximum over temperature, 26 mv full scale System noise 0.4 mvrms DNL 16-bit, no missing codes INL ±1.8 LSB CMRR 56 db (DC to 5 khz) Bandwidth 300 khz Analog Output Analog outputs 2 DAC resolution 16-bit Output range ±10 V Maximum update rate 5 ks/s simultaneous per channel, hardwaretimed AO FIFO 2,047 samples Trigger sources Software, PFI 0, PFI 1 Output current drive ±5 ma Short circuit current ±11 ma Slew rate 3 V/μs Output impedance 0.2 Ω 2 ni.com NI USB-6003 Specifications
Absolute accuracy (no load) Typical at full scale 8.6 mv Maximum over temperature, full scale 32 mv DNL 16-bit, no missing codes INL ±4 LSB Power-on state 0 V Startup glitch -7 V for 10 µs Timebase Note The following specifications apply to the sampling accuracy for hardwaretimed analog input and analog output. Timebase frequency 80 MHz Timebase accuracy ±100 ppm Timing resolution 12.5 ns Digital I/O 13 digital lines Port 0 8 lines Port 1 4 lines Port 2 1 lines Function P0.<0..7> Static digital input/output P1.0 Static digital input/output P1.1/PFI 1 Static digital input/output, counter source or digital trigger NI USB-6003 Specifications National Instruments 3
P1.<2..3> Static digital input/output P2.0/PFI 0 Static digital input/output, counter source or digital trigger Direction control Each channel individually programmable as input or output Output driver type Each channel individually programmable as open collector or active drive Absolute maximum voltage range -0.3 V to 5. 5 V with respect to D GND Pull-down resistor 47.5 kω to D GND Power-on state Input Digital Input Input voltage range (powered on) 0 to 5 V Input voltage range (powered off) 0 to 3.3 V Input voltage protection ±20 V on two lines per port (maximum of five lines for all ports) for up to 24 hours Caution Do not leave a voltage above 3.3 V connected on any DIO line for extended periods of time when the device is powered off. This may lead to long term reliability issues. Minimum V IH 2.3 V Maximum V IL 0.8 V Maximum input leakage current At 3.3 V 0.8 ma At 5 V 4.5 ma Digital Output (Active Drive) Maximum V OL (4 ma) 0.7 V Maximum V OL (1 ma) 0.2 V Minimum V OH (4 ma) 2.1 V Minimum V OH (1 ma) 2.8 V Maximum V OH 3.6 V Maximum output current per line ±4 ma 4 ni.com NI USB-6003 Specifications
Digital Output (Open Collector) Maximum V OL (4 ma) 0.8 V Maximum V OL (1 ma) 0.2 V Note Minimum V OH dependent on user-provided pull-up resistor and voltage source. Recommended pull-up resistor is 1 kω. Using a 1 kω pull-up resistor and 5 V voltage source: Minimum V OH 3.5 V Typical V OH 4.5 V Maximum output (sinking) current per line -4 ma Maximum pull-up voltage 5 V Maximum leakage current At 3.3 V 0.8 ma At 5 V 4.5 ma Counter Number of counters 1 Resolution 32-bit Counter measurements Edge counting, rising or falling Counter direction Count up Counter source PFI 0 or PFI 1 Maximum input frequency 5 MHz Minimum high pulse width 100 ns Minimum low pulse width 100 ns NI USB-6003 Specifications National Instruments 5
+5 V Power Source Output voltage +5 V, ±3% Maximum current 150 ma Overcurrent protection 200 ma Short circuit current 50 ma Overvoltage protection ±20 V Bus Interface USB specification USB Full Speed USB bus speed 12 Mb/s Physical Characteristics Dimensions Without screw terminal connector plugs 75.4 mm 86.2 mm 23.6 mm, (2.97 in. 3.40 in. 0.93 in.) With screw terminal connector plugs 93.2 mm 86.2 mm 23.6 mm, (3.67 in. 3.40 in. 0.93 in.) 6 ni.com NI USB-6003 Specifications
Figure 1. NI USB-6003 Dimensions 3.67 in (93.2 mm) 2.97 in. (75.4 mm) 0.93 in. (23.6 mm) +3 3 7 AO 3.40 in. (86.2 mm) +2 2 6 AI +0 0 4 +1 1 5 0 1 Weight Without screw terminal connector plugs 83 g (2.93 oz) With screw terminal connector plugs 105 g (3.70 oz) I/O connectors: USB Micro-B receptacle, (1) 16-position screw terminal plugs Screw-terminal wiring 1.31 to 0.08 mm 2 (16 to 28 AWG) Torque for screw terminals 0.22-0.25 N m (2.0-2.2 lb. in.) If you need to clean the module, wipe it with a dry towel. Environmental Temperature (IEC 60068-2-1 and IEC 60068-2-2) Operating 0 to 45 C Storage -40 to 85 C Humidity (IEC 60068-2-56) Operating 5 to 95% RH, noncondensing Storage 5 to 90% RH, noncondensing NI USB-6003 Specifications National Instruments 7
Pollution Degree (IEC 60664) 2 Maximum altitude 2,000 m Indoor use only. Safety This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for sensitive electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions EN 55022 (CISPR 22): Class A emissions EN 55024 (CISPR 24): Immunity AS/NZS CISPR 11: Group 1, Class A emissions AS/NZS CISPR 22: Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section. 8 ni.com NI USB-6003 Specifications
CE Compliance This product meets the essential requirements of applicable European Directives, as follows: 2006/95/EC; Low-Voltage Directive (safety) 2004/108/EC; Electromagnetic Compatibility Directive (EMC) Online Product Certification To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers This symbol indicates that waste products should be disposed of separately from municipal household waste according to WEEE Directive 2002/96/EC of the European Parliament and the Council on waste electrical and electronic equipment (WEEE). All products at the end of their life cycle must be sent to a WEEE collection and recycling center. Proper WEEE disposal reduces environmental impact and the risk to human health due to potentially hazardous substances used in such equipment. Your cooperation in proper WEEE disposal will contribute to the effective usage of natural resources. For information about the available collection and recycling scheme in a particular country, go to ni.com/environment/weee. 电 子 信 息 产 品 污 染 控 制 管 理 办 法 ( 中 国 RoHS) 中 国 客 户 National Instruments 符 合 中 国 电 子 信 息 产 品 中 限 制 使 用 某 些 有 害 物 质 指 令 (RoHS) 关 于 National Instruments 中 国 RoHS 合 规 性 信 息, 请 登 录 ni.com/environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) NI USB-6003 Specifications National Instruments 9
Device Pinout LEFT VIEW RIGHT VIEW AI GND P0.0 AI 0 P0.1 AI 4 P0.2 AI GND P0.3 AI 1 P0.4 AI 5 P0.5 AI GND P0.6 AI 2 P0.7 AI 6 P1.0 AI GND P1.1/PFI 1 AI 3 P1.2 AI 7 P1.3 AI GND P2.0/PFI 0 AO 0 D GND AO 1 + 5 V AO GND D GND Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2014 National Instruments. All rights reserved. 374372A-01 Jun14